Complete BioSafety* Systems
BioCertified**
rotors, instruments and labware contain microbiological aerosols, while
sterile filters prevent infectious materials from being expelled into
the lab environment. Together, they enhance peace of mind for users and
lab managers.
Dual System Rotor Recognition
Automatically
detects which rotor is installed and prevents the rotor from running
above its maximum-rated speed—without shutting down the run. All to
ensure the safety of your lab.
Friction Reduction System
Enables
quicker, quieter, more energy-efficient runs with better temperature
control, which improves system reliability. High-torque, switched
reluctance (SR) drive technology shortens cycle times.
Lightweight J-Lite Rotors
Consistently designed to be the most efficient rotors—for reduced run times and high g-force.
Ergonomic Design
Lower work surface for easier rotor handling. Foot pedal opens the door and improved door design swings up and out of the way.
Sustainability is Built In
Each instrument saves approximately 9.6 pounds of post-consumer materials from the landfill.
SPECIFICATIONS:
UNSPSC: | 41103905 |
Height: | 86 cm (34 in); with door open 149 cm (58.5 in) |
Width: | 71 cm (28 in) |
Depth: | 86 cm (34 in); 102 cm (40.25 in) (including air diverter extending from back panel) |
Weight: | 290 kg (640 lb.) |
Friction Reduction System (FRS): | A mechanical rotary vane pump reduces chamber pressure to 0.25 ATM (190 mm [7.5 in] Hg) |
Max Capacity Volume Range: | 6 liters |
Max Speed: | 26,000 rpm |
Maximum Heat Dissipation into Room under Steady-State Conditions: | 2.0 kW (6,900 BTU/hr) |
Noise Level 0.91 meters (3 feet) in Front of Centrifuge: | 57 dBA |
Refrigeration System: | Non-CFC, non-ozone depleting refrigerant |
Rotor ID: | Automatic |
Set Temperature: | -10° C to +40° C (in 1° C increments) |
Set Time: | 1 Min - 3 Hours, Continuous Hold |
Speed Control: | ± 10 rpm of set speed or 0.1%, whichever is greater |
Speed Range: | 100 to 26,000 rpm or equivalent RCF; 0 to 5,000 rpm for elutriation |
Temperature Control: | ± 2° C of set temperature (after equilibration) |
Zonal Capability: | Yes |
Single Tube Sample Volume Range: | 1.8 - 1,000 mL |
Ambient Temperature Range: | 16° C to 38° C |
Clearances Required: | 7.6 cm (3.0 in) sides; 16.0 cm (6.25 in) rear |
Door: | 6.1 cm (2.4 in) thick plastic resin with steel plate |
Humidity Restrictions: | < 95% (noncondensing) |
Pollution Degree: | 2 |
Rotor Chamber Diameter: | 51.3 cm (20 in) |
Time -- Actual Display: | Indicates run time remaining (timed run) or elapsed time (hold run) |
Maximum Force at r max (g): | 82,000 x g |
Acceleration: | Maximum or slow |
Deceleration: | Maximum, slow, or off |