Avanti® J-E

Avanti® J-E

MARCA: otra marca

BECKMAN COLUTER. Simple. Flexible. Control.

Consulte
Meet the Avanti J-E . . .a high powered, centrifuge solution uniquely designed to achieve a high degree of separation in a size compatible with your operations. The Avanti J-E extensive rotor library maximizes capacity and improves productivity through the use of microplates and a full complement of labware. A conveniently located touch pad designed for simplicity and ease of use allows operator selectable parameters at your fingertips.

Features

Switched reluctance (SR) drive for faster acceleration and deceleration and shorter run times
Automated Rotor Identification at low speeds with auto resets if necessary to ensure safe operations
FRS (Friction Reduction System) reduces heat generation, constant

  • Quick on the job
    Shorter run times mean more discovery for you. Avanti J-E's high torque switched reluctance (SR) drive gives you more time at speed and less time getting there.
  • Shares lab space with others The small footprint gives you the freedom to place it right in your lab, so it's where you need it, when you need it.
  • Energy Efficient Avanti J-E uses a maintenance free vacuum to reduce friction on the rotor, allowing for a smaller motor and compressor that generates less heat, and less energy usage. These energy efficient components not only prolong the life of the centrifuge, but also increase savings when amortized over time.
  • Wide Range of Rotors For Volume requirements ranging from microplates to 1-liter bottles, Avanti J-E's extensive rotor library maximizes capacity and optimizes g-force. The library includes: general purpose, high capacity (4 liter), and biosafe certified rotors with dual locking lids for safe sample transportation. Click here for the "application / rotor selection guide"
SPECIFIATIONS:
UNSPSC: 41103905
Height: 91.4 cm (36 in); with door open 152.4 cm (60 in)
Width: 63.5 cm (25 in)
Depth: 80.0 cm (31.5 in)
Weight: 267.4 kg (589 lb)
Approximate Acceleration Time: Two profiles--MAX (full acceleration to set speed) and SLOW
Approximate Deceleration Time: Three profiles--MAX, SLOW, and OFF (coast)
Electrical Requirements: 50 Hz, 180 to 253 VAC, 24 A; 60 Hz, 187to229/216to264 VAC, 24 A
Friction Reduction System (FRS): 190 mm (7.5 in) Hg
Maximum Heat Dissipation into Room under Steady-State Conditions: 2.0 kW (6,900 Btu/hr)
Noise Level 1 Meter in Front of Centrifuge: < 64 dBA
Set Speed: Actual rotor speed ± 50 rpm of set speed
Set Temperature: -10° C to +40° C in 1° C increments
Set Time: to 99 hr 59 min or continuous (hold)
Speed Range: 300 to 21,000 rpm
Temperature Control: ± 2° C of set temperature (after equilibration)
Ambient Temperature Range: 15° C to 35° C
Clearances Required: 7.6 cm (3.0 in) sides; 16.0 cm (6.25 in) rear
Humidity Restrictions: < 95% (noncondensing)
Pollution Degree: 2
Time -- Actual Display: Indicates run time remaining/displays elapsed time
Finish: Control panel: silicone elastomer. Top surface: urethane paint. Other surfaces: general purpose paint.
Availability of adapters, rotors, tubes and bottles, parts and accessories.