BECKMAN COULTER. Easy. Adaptable. Control. Experience the J6-MI centrifuge solution . . .uniquely designed to achieve key separations in volumes up to 6 Liters.
UNSPSC: | 41103905 |
Height: | 91.0 cm (35.5 in); with door open 160 cm (63 in) |
Width: | 71.1 cm (28 in) |
Depth: | 83.8 cm (33 in) |
Weight: | 252 kg (555 lb) |
Approximate Acceleration Time: | Select from 9 acceleration rates (including maximum acceleration) |
Approximate Deceleration Time: | Select from 9 deceleration rates (including maximum dynamic braking, or a coasting stop from full speed) |
g Force: | 6,835 x g |
Max Capacity Volume Range: | 6,000 mL |
Max Speed: | 6,000 rpm |
Maximum Heat Dissipation into Room under Steady-State Conditions: | 1.9 kW (7,000 Btu/hr) |
Noise Level 1 Meter in Front of Centrifuge: | < 63 dBA |
Set Speed: | Actual rotor speed ± 20 rpm of set speed |
Set Temperature: | -20° C to +40° C |
Set Time: | Up to 99 hr 59.9 min for timed runs; entries greater than 99 hr 59.9 min signal "hold" or continuous operation. |
Speed Range: | 100 to 6,000 rpm |
Temperature Control: | ± 2° for temperature from 0° to ambient, after equilibration.1 Temperature Control at higher or lower temperatures depends on the rotor and speed used (some rotors have a narrower range of Temperature Control) |
Ambient Temperature Range: | 16° C to 38° C |
Clearances Required: | 15.0 cm (6.0 in) |
Humidity Restrictions: | < 95% (noncondensing) |
Pollution Degree: | 2 |
Rotor Chamber Diameter: | 58.0 cm (23 in) |
Time -- Actual Display: | Indicates time remaining for timed runs, time elapsed for continuous operation. |
Finish: | Urethane paint on chamber door, vinyl paint on other painted surfaces, Mylar 2 on control panel. |
Operating Modes: | Key switch selects NORMAL, PROGRAM LOCK, or ZONAL operation. |