J6-MI

J6-MI

MARCA: otra marca

BECKMAN COULTER. Easy. Adaptable. Control. Experience the J6-MI centrifuge solution . . .uniquely designed to achieve key separations in volumes up to 6 Liters.

Consulte
Enjoy maintenance free operations due to the rugged, brushless induction drive which delivers high torque, fast acceleration, and extended motor life due to low wear. Experience full control with up to ten operator definable acceleration and deceleration profiles. Eliminate rotor imbalance with the exclusive ARIES rotors; designed to auto correct opposing, imbalanced loads up to 100 grams without manual intervention or delay.

FEATURES:
The versatile J6-MI delivers a high level of performance in separation techniques for bioprocessing, clinical, research and blood separation laboratories.
  • Adaptable design - no maintenance, worry free performance
  • Self balancing operations - fast, easy and safe
  • Energy efficient - less noise for a comfortable work environment
  • Total control - operator selectable profiles adapt to your lab procedures today and in the future . . .
 
SPECIFICATIONS:
UNSPSC: 41103905
Height: 91.0 cm (35.5 in); with door open 160 cm (63 in)
Width: 71.1 cm (28 in)
Depth: 83.8 cm (33 in)
Weight: 252 kg (555 lb)
Approximate Acceleration Time: Select from 9 acceleration rates (including maximum acceleration)
Approximate Deceleration Time: Select from 9 deceleration rates (including maximum dynamic braking, or a coasting stop from full speed)
g Force: 6,835 x g
Max Capacity Volume Range: 6,000 mL
Max Speed: 6,000 rpm
Maximum Heat Dissipation into Room under Steady-State Conditions: 1.9 kW (7,000 Btu/hr)
Noise Level 1 Meter in Front of Centrifuge: < 63 dBA
Set Speed: Actual rotor speed ± 20 rpm of set speed
Set Temperature: -20° C to +40° C
Set Time: Up to 99 hr 59.9 min for timed runs; entries greater than 99 hr 59.9 min signal "hold" or continuous operation.
Speed Range: 100 to 6,000 rpm
Temperature Control: ± 2° for temperature from 0° to ambient, after equilibration.1 Temperature Control at higher or lower temperatures depends on the rotor and speed used (some rotors have a narrower range of Temperature Control)
Ambient Temperature Range: 16° C to 38° C
Clearances Required: 15.0 cm (6.0 in)
Humidity Restrictions: < 95% (noncondensing)
Pollution Degree: 2
Rotor Chamber Diameter: 58.0 cm (23 in)
Time -- Actual Display: Indicates time remaining for timed runs, time elapsed for continuous operation.
Finish: Urethane paint on chamber door, vinyl paint on other painted surfaces, Mylar 2 on control panel.
Operating Modes: Key switch selects NORMAL, PROGRAM LOCK, or ZONAL operation.


Availability of adapters, rotors, tubes and bottles, parts and accessories.